HiPP-Lab XPS with Liquid/Gas Reaction Cell

[ Available Winter 2025 ]
HiPP-Lab XPS Setup

The Scienta Omicron HiPP Lab is a state-of-the-art ambient-pressure photoelectron spectroscopy (AP-PES) system with high-temperature stage designed for XPS in gas and/or liquid environments and capable of operando electrochemistry / catalysis measurements. A couple gas reaction cell and preparation chamber, with attached glovebox, allow preparation and processing of samples without exposure to atmosphere. 

Scienta Omicron HiPP Lab System:

  •  AP-PES system for XPS measurements at pressures to 25 mbar
  • Guaranteed 0.6 eV energy resolution on Ag3d5/2 (assuming a HiPPLab with a standard chamber)
  • 20 μm spatial resolution along the analyser slit
  • μ-metal chamber for reliable shielding properties ensuring high transmission
  • Sample alignment camera 
  • Automated Stage 
  • Electrochemistry kit for HiPP Lab system 
  • Preparation module for HiPP system with 4-axis sample manipulator and heating stage 
  • Residual Gas Analyser, 1-200 amu 
  • Automated gas flow management system
  • Plasma source package for chamber cleaning
  • VUV Light Source for UPS Measurement
  • Preparation module with Gas Reaction Cell
  • Ion sputter gun (0.12-5keV; adjustable spot size 3-20 mm dia.; Operates with backfilled Ar/O/H)
  • Glovebox for the preparation of samples without atmospheric exposure

Analyzer:

- Parallel image mode capabilities for studying on-sample variations
- Swift acceleration mode for fast acquisition
- Extremely stable and low noise electronics
- Strictly non-magnetic materials used in the interior
- Acquisition modes: Swept, fixed, imaging
- CCD/MCP detector with Pulsed and ADC mode operation

Energy resolution: < 5 meV FWHM at 5 eV pass energy and 10 eV kinetic energy 
- Pass energies: 2, 5, 10, 20 eV
- Transmission mode kinetic energy range: 5 eV – 20 eV
- Angular mode kinetic energy range: 5 eV – 100 eV
- Spatial mode kinetic energy range: 5 eV – 20 eV - Energy resolution: < 40 meV FWHM at 20 eV pass energy and 6000 eV kinetic energy
- Transmission mode kinetic energy range: 0 eV – 6000 eV

X-ray Source:

- High energy-resolution monochromated Al-Ka X-ray source
- 500 mm Rowland circle diameter
- Small X-ray spot (Typically 80W on 300μm Spot) with high flux density, optimised for operation in APPES experiments

 VUV Light Source:

  • He I photon flux up to 6x10^12 photons/sec or 1x10^16 photons/(s*sr)
  • Source is suitable for excitation of different gases (H, He, Ne, Ar, Kr, Xe)

Sample Stage:

- 4 axes (x/y/z, polar rotation) analysis module manipulator
- Scienta Omicron's flag-style sample environment
- Electrically isolated from ground
- Non-magnetic reference thermocouples for measurement of sample temperature 
- IR laser sample heating with pyrometer surface temperature measurement - Fiber-coupled diode laser, wavelength 915nm, 200W laser power
- Target temperature: RT - 650 °C

Preparation Chamber with High-Pressure Gas Reaction Cell:

- Radiative sample heating with a compact, rugged wire heater, non-reactive to oxidizing or reducing gases
- 1300 K at up to 10 bar static gas pressure (depending on gas species), 1070 K at gas flow conditions with 0.5 bar pressure difference between in- and outlet
- Heating rate up to 200 K/min
- Type S thermocouple located close to the sample acceptance for accurate temperature reading
- The heater cell is shut off with a Viton-sealed door with a latch mechanism actuated by a wobblestick and a heavy-duty linear drive mechanism
- Heater cell and door are water-cooled
- The heater cell is housed in a UHV chamber 
- Pressure Control 
- Gas Sampling
 

Rates

UVA RATES:

AP-PES Instrument Usage: $20/hr

AP-PES Operator Costs:  $60/hr

Prep/Gas Reaction Chamber Usage: = $15/day

Electrochemistry Cell Usage:  = $15/day

EDUCATIONAL, GOVERNMENT & NON-PROFIT INSTITUTION RATES :

AP-PES Instrument Usage: $60/hr

AP-PES Operator Costs:  $70/hr

Prep/Gas Reaction Chamber Usage: = $30/day

Electrochemistry Cell Usage:  = $30/day

INDUSTRIAL PARTNER RATES:

AP-PES Instrument Usage:  $150/hr

AP-PES Operator Costs:  $125/hr

Prep/Gas Reaction Chamber Usage:  = $60/day.

Electrochemistry Cell Usage:  = $60/day